The AT47AV series has been upgraded to the AT47AV series, as an iterative upgrade of the same series. The instrument is controlled by a high-performance ARM9 microprocessor with real-time high-speed data acquisition. The newly upgraded full mirror touch panel uses Onmeas' 9th generation ATOS operating system, supporting both Chinese and English interfaces, making operation intuitive, easier and smarter.
The newly designed AT47AV series comes standard with 64G large internal memory, communication mapped as USB drive. Using real-time FPGA circuitry, it breaks through the sampling rate limitation. Combined with efficient, high-precision analog circuits, it achieves high precision and ultra-high-speed sampling, with temperature test resolution improved to 0.01°C. The instrument features channel isolation voltage of 350VDC, 230VAC (full isolation between digital and analog signals), suitable for complex systems requiring simultaneous temperature and voltage monitoring, such as new energy and automation control fields.
The instrument has built-in LAN interface, RS232, USB-232, VCOM virtual serial port. RS-232 and USB-232 interfaces support Modbus(RTU) protocol, and the LAN port supports Modbus(TCP) protocol. The instrument can be controlled and data can be collected through a computer, effectively used for remote control and data collection and analysis.
The instrument supports USB communication, and data collection, analysis and printing can be realized through the standard computer software. It supports USB disk storage for real-time storage of sampled data. Users can independently calibrate each channel.
| Parameter | AT4740AV | |
|---|---|---|
| Basic Parameters | ||
| Number of Channels | 40 channels | |
| Thermocouple Type | Thermocouple: J/K/T/E/S/N/B/R type; RTD: PT100, PT1000, CU50; Humidity: Humidity test module (optional); Voltage: DCV | |
| Resolution | Temperature: 0.01℃; DCV: 100mV~100V | |
| Scan Speed | Fast: 500ms/channel; Slow: 1s/channel | |
| Calibration | Error correction per channel | |
| Test Terminal Voltage | Channel isolation voltage DC 350V; AC 230V | |
| Test Range | ±0.2%±2 digits (excluding thermocouple); PT100: ±0.6℃, PT1000: ±0.6℃, CU50: ±0.6℃ | |
| Basic Accuracy | K-type thermocouple -100℃ to 0℃ ±1.2℃, 0℃ to 1350℃ ±0.8℃ (varies by thermocouple type, refer to user manual for details) | |
| Cold Junction Compensation | Accuracy 0.5℃ | |
| Other Functions | Over-limit and under-limit alarm, configurable per channel; Data recording function; USB storage support | |
| General Parameters | ||
| Dimensions | Instrument Dimensions | 294(W) x 130(H) x 316(D) mm |
| Package Dimensions | 350mm x 200mm x 400mm | |
| Weight | Net Weight | 3kg |
| Gross Weight | 4.5kg | |
| Power Requirements | Voltage: 100VAC~240VAC, Frequency: 47.5-52.5Hz, Power: 10VA | |
| Interface | RS232C interface, USB drive interface, Type-C communication interface, LAN interface | |
| Other | 7-inch TFT-LCD true color display, Open thermocouple detection function; Standard software: ATS45 data acquisition software; Accessories: 2m thermocouple, 8 pcs/set, Type-C communication cable | |