The ATH9006 Comprehensive Safety Tester is a multi-functional, high-performance, high-reliability electrical safety testing instrument, integrating seven test functions: AC withstanding voltage, DC withstanding voltage, insulation resistance, grounding resistance, leakage current, AC power, and startup test. It features 1.5% measurement accuracy with simultaneous display of multiple parameters. The system uses an ARM processor with a 10-inch HD TFT touchscreen for fast and accurate parameter measurement. The instrument is equipped with Handler, RS-232, and other interfaces to meet various testing communication requirements.
The ATH9006 Comprehensive Safety Tester is primarily designed to provide safety testing solutions for the electric vehicle, power grid, photovoltaic, home appliance, information equipment, and medical equipment industries, as well as innovative solutions for various electrical measurements.
| Parameter | ATH9006 | |
|---|---|---|
| Basic Information | ||
| Test Functions | AC/DC Withstanding Voltage, Insulation Resistance, Grounding Resistance, Leakage Current, AC Power, Startup (optional) | |
| Interfaces | RS-232/RS-485/USB/Handler; External Remote Control / Alarm Light Interface | |
| Standard Configuration | Tri-color alarm light; Remote switch with external start/stop signal support | |
| Protection Functions | Over-voltage, Over-current | |
| Other Functions | 9" Color LCD | |
| AC Withstand Voltage | ||
| Rated Output | 5kV AC, 100mA | |
| Actual Output Power | 500VA | |
| Output Voltage | 100 to 5000VAC (Resolution: 1V) | |
| Voltage Accuracy | ±(1.5% of set value + 3V) | |
| Output Frequency | 50Hz / 60Hz | |
| Measurement Current | 0.002 to 9.999mA / 100.00mA (Resolution: 0.001mA / 0.01mA) | |
| Current Accuracy | 1.5% × Reading + 0.01mA | |
| Voltage Rise Time | 1s to 999s | |
| Test Time | 1s to 999s | |
| Voltage Fall Time | 1s to 999s | |
| DC Withstand Voltage | ||
| Rated Output | 6kV DC / 10mA | |
| Actual Output Power | 60VA | |
| Output Voltage | 100V to 6kV DC (Resolution: 1V) | |
| Voltage Accuracy | ±(1.5% of set value + 3V) | |
| Current Accuracy | 2% × Reading + 0.01mA | |
| Voltage Rise Time | 1s to 999s | |
| Test Time | 1s to 999s | |
| Voltage Fall Time | 1s to 999s | |
| Insulation Resistance | ||
| Rated Output | 1kV / 10GΩ | |
| Output Voltage | 100V to 1000V (Resolution: 1V) | |
| Voltage Accuracy | ±(1.5% of set value + 3V) | |
| Measurement Range | 0.1MΩ to 10GΩ | |
| Resistance Accuracy | 100-499V: 1MΩ to 16GΩ ±(5% Reading + 0.5MΩ), 1GΩ to 10GΩ for reference only; 500-1000V: <100MΩ to 1GΩ ±(3% Reading + 2MΩ), 1GΩ to 10GΩ ±(10% Reading + 5MΩ) | |
| Test Time | 1s to 999s | |
| Grounding Resistance | ||
| Rated Output | 2.0 to 40A, 10.0 to 600mΩ | |
| Output Current | 40A | |
| Output Voltage | <12V | |
| Frequency | 50Hz / 60Hz | |
| Test Time | 1s to 999s | |
| Leakage Current | ||
| Test Voltage Output | Single-phase (requires external isolation transformer and variable frequency power supply) | |
| Test Type | G-L, G-N, AUTO (G-L, G-N) | |
| AC Voltage Measurement | 60.0V to 300.0V (Resolution: 0.1V, Accuracy: 1.5% × Reading + 3V) | |
| AC Current Measurement | 0.000mA to 20.00mA (Resolution: 0.001mA / 0.01mA, Accuracy: ±(1.5% Reading + 5 digits)) | |
| Test Time | 1s to 999s | |
| AC Power | ||
| Test Voltage Output | Single-phase (requires external isolation transformer and variable frequency power supply) | |
| AC Voltage Measurement | 60.0V to 300.0V (Resolution: 0.1V, Accuracy: ±(0.25% × Reading + 0.25% × Range)) | |
| AC Current Measurement | 0 to 30A (Resolution: 0.001A / 0.01A, Accuracy: ±(0.25% × Reading + 0.25% × Range)) | |
| Active Power Measurement | 0 to 6000W (Resolution: 0.1W / 1W, Accuracy: ±(0.25% × Reading + 0.25% × Range)) | |
| Test Time | 1s to 999s | |
| Startup Function | ||
| AC Voltage Measurement | 60.0V to 300.0V (Resolution: 0.1V, Accuracy: ±(0.25% Reading + 0.25% Range)) | |
| AC Current Measurement | 0 to 30A | |
| Current Setting Range | Upper limit: 0.001 to 30.00A, Lower limit: 0.000 to 30.00A | |
| General Parameters | ||
| Dimensions | Instrument Dimensions | 470 × 133 × 491 mm |
| Package Dimensions | - | |
| Weight | Net Weight | 30 kg |
| Gross Weight | - | |
| Power Requirements | 220V to 240V AC, 47.5 to 52.5Hz, 300VA | |
| Interfaces | RS-232/RS-485/USB/Handler; External Remote Control / Alarm Light Interface | |
| Others | 9" Color LCD | |